1 ABL5 - Troubleshooting 1 RED SYSTEM
2 ABL5 - Troubleshooting 2 RED SYSTEM System Status RecordsCal Status Operators Manual, Chapter 8 Troubleshooting Tools
3 ABL5 - Troubleshooting 3 RED SYSTEM System Status All error messages are registered in the System Status Messages will be cleared when appropriate conditions have been met Divided into 3 Groups General record: mechanical or electronic malfunction messages Measurement record: errors recorded during measurements Calibration record: errors recorded during calibrations
4 ABL5 - Troubleshooting 4 RED SYSTEM Records ? in System Status field indicates records present in: General record Calibration record No ? from records present in: Measurement record System Status field
5 ABL5 - Troubleshooting 5 RED SYSTEM General Record Refers to mechanical or electronic malfunction messages lack of liquids or gases temperature faults during measurement or calibration mechanical defects or failures 10. Not used 11. MAINT. PROG ERROR 12. Not used 13. Not used 14. Not used 15. LS ERROR 16. PRINTING ERROR 17. SENDING ERROR 18. LEAK DETECTED 1. TEMP ERROR 2. GAS ERROR 3. RINSE ERROR 4. BARO ERROR 5. Not used 6. CAL NEEDED 7. Not used 8. Not used 9. pO2 FLUSH ERROR
6 ABL5 - Troubleshooting 6 RED SYSTEM Messages Messages in System Status RINSE ERROR Stays in START UP Records indicated by a flashing ? General record Calibration record Records without screen message Measurement record
7 ABL5 - Troubleshooting 7 RED SYSTEM Marking of Results Unreliable results due to an error will be noted with questions marks,"?" "?" for all the relevant measured and derived parameters
8 ABL5 - Troubleshooting 8 RED SYSTEM System status
9 ABL5 - Troubleshooting 9 RED SYSTEM General recordMeasurement recordCalibration record 1. TEMP ERROR1. pH ERROR1. pH ERROR 2. GAS ERROR2. pCO 2 ERROR2. pCO 2 ERROR 3. RINSE ERROR3. pO 2 ERROR3. pO 2 ERROR 4. BARO ERROR4. BARO OUT OF RANGE4. BARO OUT OF LIMITS 5. Not used5. Not used5. CAL 2 ERROR 6. CAL NEEDED6. Not used6. Not used 7. Not used7. MEAS. TEMP ERROR7. CAL TEMP ERROR 8. Not used8. Not used8. Not used 9. pO2 FLUSH ERROR9. Not used9. Not used 10. Not used10. MEAS. LS ERROR10. Not used 11. MAINT. PROG ERROR11. MEAS. ERROR11. Not used 12. Not used12. CAL RINSE ERROR 13. Not used13. CAL SOLUTION ERROR 14. Not used 15. LS ERROR 16. PRINTING ERROR 17. SENDING ERROR 18. LEAK DETECTED