1 ABL5 - Calibration 1 RED SYSTEM ± Sensitivity ± Zero point ± Drift ± 1 point calibration ± 2 point calibration ± Status value
2 ABL5 - Calibration 2 RED SYSTEM What is Sensitivity It is the slope of the calibration line B C Cal 2 Cal 1 Electrode signal pH
3 ABL5 - Calibration 3 RED SYSTEM What is Status It is the vertical position of the calibration line pH B C Cal 2 Cal 1 Theoretical pH electrode Real pH electrode Electrode signal Status
4 ABL5 - Calibration 4 RED SYSTEM What Is Drift Basically it is the new calibration minus the old calibration New (7.383) - Old (7.383) x Sens. New (6.841) - Old (6.841) x Sens. High Low
5 ABL5 - Calibration 5 RED SYSTEM What is Zero point Current running when measuring on Gas 2 (O 2 =0%)
6 ABL5 - Calibration 6 RED SYSTEM 1 point Calibration Calibrates all electrodes at 1 point pH ? C Cal 1 Electrode signal
7 ABL5 - Calibration 7 RED SYSTEM 2 point Calibration Calibrates all electrodes at 2 points pH B C Cal 1 Cal 2 Electrode signal