1 MEK-6400 series Technical seminar Sensors adjustment
2 Adjustment of the sensors MEK-6400 series have 1 HGB measuring sensor, and 6 liquid detecting sensors. HGB measuring sensor WBC manometer sensor ( High, Low ) RBC manometer sensor ( High, Low ) Diluent and detergent detecting sensor Lysing reagent detecting sensor This section is how to adjust these sensors.
3 Adjustment on screen Sensor voltage adjustment is available on the screen. (*)except HGB sensor voltage
4 HGB measuring sensor HGB Measuring sensor
5 Adjustment of HGB measuring sensors voltage There is a HGB sensor with a WBC measurement bath in measurement unit MC-640V. It is necessary to keep measurement reliability of HGB, the HGB sensor voltage should be adjusted around 3.0V in a state of HGB LED on. The sensor voltage has to be less than 0.5V in a state of HGB LED off. When the HGB sensor voltage is abnormal, * mark is displayed if the voltage is less than 0.5V when HGB LED is off. ! mark is displayed if the voltage is more than 4.5V when HGB LED is on. ? mark is displayed if the voltage is less than 1.5V when HGB LED is on. In such a case, the reliability of HGB results becomes lower.
6 Adjustment of HGB measuring sensors voltage Adjustment method of the HGB sensor in the following. Choose OTHER key from MENU screen. After that, choose SENSOR MONITOR function, so that you can see all sensors voltage on the screen.
7 Adjustment of HGB measuring sensors voltage Remove the side panel. Check the WBC measurement bath is filled by diluent. Turn the variable resistance of the HGB sensor voltage on UT-7201 ( HGB AMP BD, cf. illustration ) and adjust it so that the HGB sensor voltage becomes about 3.0V. HGB sensor VR
8 Adjustment of HGB measuring sensors voltage Check HGB sensor voltage in a state of LED On is about 3.0V ( this figure is bad case ) Check HGB sensor voltage in a state of LED Off is less than 0.5V After adjustment, replace the side panel.
9 WBC/RBC Manometer sensor WBC manometer (High) WBC manometer (Low) RBC manometer (Low) RBC manometer (High)
10 Adjustment of WBC/RBC Manometer sensors voltage Two manometers for WBC and RBC absorb diluted sample of 360uL from apertures in measurement unit MC-640V. Each manometer has two liquid surface detecting sensors, higher and lower sensor. In the state that manometer is filled by diluent, the manometers sensors voltage should be lower than 1.5V. In the state that manometer is empty, the manometers sensors voltage should be higher than 3.5V A device judges to be or not to be of water at 2.5V. But if sensor voltages are between 2.25V V, device cannot detect liquid surface precisely. To keep measurement reliability, device displays alarm such as Manometer dirty
11 Adjustment of WBC/RBC Manometer sensors voltage To adjust the manometer voltages, you have to change the operator to Factory Choose Setting in MENU, and select USERS. Choose "Factory" and push CHANGE" key to change the operator. You have to enter the password. Enter "4321 as a password of "Factory".
12 Adjustment of WBC/RBC Manometer sensors voltage Adjustment method of the WBC/RBC Manometer sensors in the following. Choose OTHER key from MENU screen. After that, choose SERVICE function. Choose MONITOR key form SERVICE screen, after that choose Sensor monitor function.
13 Adjustment of WBC/RBC Manometer sensors voltage Different point of two Sensor monitor screen, in normal menu and in service menu, is INPUT CAL" key in service menu. To adjust the manometer sensor voltages, push "INPUT CAL" key and change coefficient factors.
14 Adjustment of WBC/RBC Manometer sensors voltage Displayed contents are as follows. WBC up: Upper sensor coefficient of WBC manometer WBC low: Lower sensor coefficient of WBC manometer RBC up: Upper sensor coefficient of RBC manometer RBC low: Lower sensor coefficient of RBC manometer
15 Adjustment of WBC/RBC Manometer sensors voltage Move a cursor with arrow key and put ENTER key after the input with numerical value. Push OK key to back to Sensor monitor screen, and check that the sensor voltage of manometers in filled by diluent is lower than 1.5V And check that the sensor voltage is adjusted more than 3.5V in the state that there is no diluent in the manometer. Repeat above operations till these voltages become reasonable values.
16 Diluent /Detergent sensor, Lysing reagent sensor Lysing reagent sensor Diluent / Detergent sensor
17 Adjustment of Diluent /Detergent, Lysing reagent sensor voltage MEK-6400 has two reagent detection sensors which for diluent / detergent detection, and lysing reagent detection in Inlet and outlet unit JQ-640V. Each liquid sensor voltage is less than 1.5V in a state to filled by reagent, and it is necessary to be adjusted than 3.5V in a empty state. A device judges to be or not to be of water at 2.5V. But if sensor voltages are between 2.25V V, device cannot detect liquid surface precisely.
18 Adjustment of Diluent /Detergent, Lysing reagent sensor voltage Displayed contents are as follows. Diluent: Diluent and Detergent detecting sensor coefficient. Lysing: Lysing reagent detecting sensor coefficient of WBC manometer
19 Adjustment of Diluent /Detergent, Lysing reagent sensor voltage Move a cursor with arrow key and put ENTER key after the input with numerical value. Push OK key to back to Sensor monitor screen, and check that the sensor voltage in filled by reagent is lower than 1.5V And check that the sensor voltage is adjusted more than 3.5V in the state that there is no diluent. Repeat above operations till these voltages become reasonable values.