ABL5 - Maintenance 1 RED SYSTEM
ABL5 - Maintenance 2 RED SYSTEM Easy Maintenance Divided into 4 groups Daily Weekly Monthly Half yearly Maintenance programs available Rinse Cleansing Protein removal Hold analyzer Refill LS adjustment
ABL5 - Maintenance 3 RED SYSTEM Maintenance Programs ABL5 Main Menu CalibrationUtilities System status Maintenance 1. Rinse 2. Cleansing 3. Protein removal 4. Hold analyzer 5. Refill 6. LS adjustment Print/Send
ABL5 - Maintenance 4 RED SYSTEM Daily Maintenance Check solution level Examine / clean inlet Check level in waste container Perform cleaning program use S5332 Cleaning Solution
ABL5 - Maintenance 5 RED SYSTEM Weekly Maintenance Perform Protein removal program use S5362 Hypochlorite Solution
ABL5 - Maintenance 6 RED SYSTEM Monthly Maintenance Remembrane electrodes
ABL5 - Maintenance 7 RED SYSTEM Half Yearly Maintenance Replace Pump Tubing
ABL5 - Maintenance 8 RED SYSTEM Replacement of Electrodes Recommended pO 2 every 24 months pH and pCO 2 every 18 months Reference electrode every 18 months Or when needed
ABL5 - Maintenance 9 RED SYSTEM Warranty policy ( between RADIOMETER Medical ApS, ISD and the Distributor) Instruments12 months from date of installation, max. 18 months from date of invoice Electrodes2 years from date of invoice* or expiry date (18 months after production)** - min. 12 months from date of invoice) AccessoriesExpiry date Spare parts12 months from date of use, max. 4 years from date of invoice Factory repairs 6 months from date of invoice * ABL5XX/ABL6XX/ABL7XX, excluding metabolite ** If expiry code is marked with an asterisk: extension of expiry date as outlined on Codes of Expiry Dates.
ABL5 - Maintenance 10 RED SYSTEM Trade-in Parts included:Major PCBs and a few mechanical parts Returned parts are refurbished and sold with same warranty as new parts