ABL5 - Introduction 1 RED SYSTEM ABL5 pH, pCO 2,pO 2Blood Gas:pH, pCO 2, pO 2 Sample volume 85 L for all three parameters Sample volume 35 L for pH only Measuring time 60 seconds for all three parameters 45 seconds for pH only Automatic calibration schedule Standby option to reduce operating costs in longer periods of non-use Low maintenance
ABL5 - Introduction 2 RED SYSTEM ABL5 Aspiration probe can be positioned for both syringe and capillary samples Safe waste system a safety bar protects you from the probe sealable waste container for safe disposal Identification input Patient ID, User ID Sample type QC type and level Input Data Patient temperature, FIO 2 ctHb, QC ampoule temperature
ABL5 - Introduction 3 RED SYSTEM What is What - Front Thermal Printer Display Keypad Aspirate Key Front Cover Aspiration Inlet Rinse Solution Calibrating Solution 1 Calibrating Solution 2 Waste Container
ABL5 - Introduction 4 RED SYSTEM What is What - Wet Section Pump pO 2 Electrode pCO 2 Electrode Preheater pH Electrode Reference Electrode Measuring Chamber Window Membrane Valve Unit
ABL5 - Introduction 5 RED SYSTEM What is What - Rear 3 Ports Line Voltage Selector (110, 120, 230, 240 V) Power Switch Power Socket HIGH GAS Nozzle (11.2 % CO 2 ) LOW GAS Nozzle (5.6 % CO 2 ) Ground Terminal
ABL5 - Introduction 6 RED SYSTEM What is What - Gas Supply Main Valve Reduction Valves Gas Cylinder 1 Gas Cylinder 2 Tubings
ABL5 - Introduction 7 RED SYSTEM Numeric Keypad Paper Feed Button Numerical Keys Decimal Point Key ENTER Key BACKSPACE Key Aspirate Key
ABL5 - Introduction 8 RED SYSTEM What is What - Display Instrument Status Field System Status Field Data Field Softkey Label Fields Softkeys
ABL5 - Introduction 9 RED SYSTEM Wet Section Diagram