MEK-6400 series Technical seminar Sensors adjustment.

Презентация:



Advertisements
Похожие презентации
MEK-6400 series Technical seminar ALARM & SYSTEM ERROR.
Advertisements

Presentation about hobby that would be my favourite Surfing made by Ezjaev A.V. made by Ezjaev A.V.
PAT312, Section 21, December 2006 S21-1 Copyright 2007 MSC.Software Corporation SECTION 21 GROUPS.
Using Dreamweaver MX Slide 1 Window menu Manage Sites… Window menu Manage Sites… 2 2 Open Dreamweaver 1 1 Set up a website folder (1). Click New…
SJB (Smart Junction Box) Copyright 2009 All rights reserved. No part of this material may be reproduced, stored in any retrieval system or transmitted.
SPLAY TREE The basic idea of the splay tree is that every time a node is accessed, it is pushed to the root by a series of tree rotations. This series.
USB Download Manual (v1.3) (GP2 Year 2010) LG Electronics/ LCD TV Division Feb. 17 th, Applied Models & Notice - File Copy - User Download Mode.
S4-1 PAT328, Section 4, September 2004 Copyright 2004 MSC.Software Corporation SECTION 4 FIELD IMPORT AND EXPORT.
Taking out Money from a Cash Machine Authors: Aleksey Ermolaev, Daria Zaitseva, Maria Leontyeva, Anatoly Leshchev, Form 10 pupils Teacher: V. V. Sergoushina,
The Law of Demand The work was done by Daria Beloglazova.
Sequences Sequences are patterns. Each pattern or number in a sequence is called a term. The number at the start is called the first term. The term-to-term.
ADVANCED DRESS-UP FEATURES 39. Once OK has been selected, your part will appear with the filleted area highlighted by orange lines at the boundaries.
Project My Future Profession Made by: Kolesnikova Olga Form: 8 «V»
Correlation. In statistics, dependence refers to any statistical relationship between two random variables or two sets of data. Correlation refers to.
Time-Series Analysis and Forecasting – Part IV To read at home.
11 BASIC DRESS-UP FEATURES. LESSON II : DRESS UP FEATURES 12.
S5-1 PAT328, Section 5, September 2004 Copyright 2004 MSC.Software Corporation SECTION 5 RESULTS TITLE EDITOR.
Our parents it is all in our life. They know more us, they us learn us how correctly to live that is possible, and that isn't present.
Combination. In mathematics a combination is a way of selecting several things out of a larger group, where (unlike permutations) order does not matter.
Business Statistics 1-1 Chapter Two Describing Data: Frequency Distributions and Graphic Presentation GOALS When you have completed this chapter, you will.
Транксрипт:

MEK-6400 series Technical seminar Sensors adjustment

Adjustment of the sensors MEK-6400 series have 1 HGB measuring sensor, and 6 liquid detecting sensors. HGB measuring sensor WBC manometer sensor ( High, Low ) RBC manometer sensor ( High, Low ) Diluent and detergent detecting sensor Lysing reagent detecting sensor This section is how to adjust these sensors.

Adjustment on screen Sensor voltage adjustment is available on the screen. (*)except HGB sensor voltage

HGB measuring sensor HGB Measuring sensor

Adjustment of HGB measuring sensors voltage There is a HGB sensor with a WBC measurement bath in measurement unit MC-640V. It is necessary to keep measurement reliability of HGB, the HGB sensor voltage should be adjusted around 3.0V in a state of HGB LED on. The sensor voltage has to be less than 0.5V in a state of HGB LED off. When the HGB sensor voltage is abnormal, * mark is displayed if the voltage is less than 0.5V when HGB LED is off. ! mark is displayed if the voltage is more than 4.5V when HGB LED is on. ? mark is displayed if the voltage is less than 1.5V when HGB LED is on. In such a case, the reliability of HGB results becomes lower.

Adjustment of HGB measuring sensors voltage Adjustment method of the HGB sensor in the following. Choose OTHER key from MENU screen. After that, choose SENSOR MONITOR function, so that you can see all sensors voltage on the screen.

Adjustment of HGB measuring sensors voltage Remove the side panel. Check the WBC measurement bath is filled by diluent. Turn the variable resistance of the HGB sensor voltage on UT-7201 ( HGB AMP BD, cf. illustration ) and adjust it so that the HGB sensor voltage becomes about 3.0V. HGB sensor VR

Adjustment of HGB measuring sensors voltage Check HGB sensor voltage in a state of LED On is about 3.0V ( this figure is bad case ) Check HGB sensor voltage in a state of LED Off is less than 0.5V After adjustment, replace the side panel.

WBC/RBC Manometer sensor WBC manometer (High) WBC manometer (Low) RBC manometer (Low) RBC manometer (High)

Adjustment of WBC/RBC Manometer sensors voltage Two manometers for WBC and RBC absorb diluted sample of 360uL from apertures in measurement unit MC-640V. Each manometer has two liquid surface detecting sensors, higher and lower sensor. In the state that manometer is filled by diluent, the manometers sensors voltage should be lower than 1.5V. In the state that manometer is empty, the manometers sensors voltage should be higher than 3.5V A device judges to be or not to be of water at 2.5V. But if sensor voltages are between 2.25V V, device cannot detect liquid surface precisely. To keep measurement reliability, device displays alarm such as Manometer dirty

Adjustment of WBC/RBC Manometer sensors voltage To adjust the manometer voltages, you have to change the operator to Factory Choose Setting in MENU, and select USERS. Choose "Factory" and push CHANGE" key to change the operator. You have to enter the password. Enter "4321 as a password of "Factory".

Adjustment of WBC/RBC Manometer sensors voltage Adjustment method of the WBC/RBC Manometer sensors in the following. Choose OTHER key from MENU screen. After that, choose SERVICE function. Choose MONITOR key form SERVICE screen, after that choose Sensor monitor function.

Adjustment of WBC/RBC Manometer sensors voltage Different point of two Sensor monitor screen, in normal menu and in service menu, is INPUT CAL" key in service menu. To adjust the manometer sensor voltages, push "INPUT CAL" key and change coefficient factors.

Adjustment of WBC/RBC Manometer sensors voltage Displayed contents are as follows. WBC up: Upper sensor coefficient of WBC manometer WBC low: Lower sensor coefficient of WBC manometer RBC up: Upper sensor coefficient of RBC manometer RBC low: Lower sensor coefficient of RBC manometer

Adjustment of WBC/RBC Manometer sensors voltage Move a cursor with arrow key and put ENTER key after the input with numerical value. Push OK key to back to Sensor monitor screen, and check that the sensor voltage of manometers in filled by diluent is lower than 1.5V And check that the sensor voltage is adjusted more than 3.5V in the state that there is no diluent in the manometer. Repeat above operations till these voltages become reasonable values.

Diluent /Detergent sensor, Lysing reagent sensor Lysing reagent sensor Diluent / Detergent sensor

Adjustment of Diluent /Detergent, Lysing reagent sensor voltage MEK-6400 has two reagent detection sensors which for diluent / detergent detection, and lysing reagent detection in Inlet and outlet unit JQ-640V. Each liquid sensor voltage is less than 1.5V in a state to filled by reagent, and it is necessary to be adjusted than 3.5V in a empty state. A device judges to be or not to be of water at 2.5V. But if sensor voltages are between 2.25V V, device cannot detect liquid surface precisely.

Adjustment of Diluent /Detergent, Lysing reagent sensor voltage Displayed contents are as follows. Diluent: Diluent and Detergent detecting sensor coefficient. Lysing: Lysing reagent detecting sensor coefficient of WBC manometer

Adjustment of Diluent /Detergent, Lysing reagent sensor voltage Move a cursor with arrow key and put ENTER key after the input with numerical value. Push OK key to back to Sensor monitor screen, and check that the sensor voltage in filled by reagent is lower than 1.5V And check that the sensor voltage is adjusted more than 3.5V in the state that there is no diluent. Repeat above operations till these voltages become reasonable values.